How do I change the deuterium lamp on an Agilent 1100/1200 series HPLC system?

LD-AGI-104LLIf the lamp on your Agilent 1100 or 1200 series HPLC system fails or if the baseline noise / drift exceeds your application limits, follow the instructions below to replace your Deuterium lamp.
  1. Switch off your system and allow to cool.
  2. Remove the front cover.
  3. Disconnect the electrical connection and twist the lamp to unscrew it from the DAD.
  4. Remove your new lamp from the packaging, taking care not to touch the glass, and twist it into place in the DAD.
  5. Tighten the screws and reconnect the lamp to the electrical connection.
  6. Replace the front cover.
  7. Reset the lamp counter and turn the lamp on - The new 'C' type detectors use chipped lamps which automatically count used lamp time.
  8. Allow the lamp to warm up for 10-20 minutes before use.
  Order your replacement lamp Kinesis have been manufacturing lamps for over 10 years now and with sales of over 10,000 lamps per year, we are the largest 3rd party lamp manufacturer in the world. This means that Kinesis are able to offer the highest quality lamps with large price saving benefits. The benefits of Kinesis lamps include:
  • Large Price Savings over branded spares with no reduction in quality
  • Bare lamps sourced from the same factories as used by the original instrument manufacturers
  • All guarantees and warranties match those offered by the OEM
  • ISO certified and industry verified manufacturing processes
  • Lamps individually QC tested at every processing step
Select your instrument model below to buy online:
ABI Agilent Aminco Analytik Jena Bausch & Lomb
Beckman Biorad Bischoff Bruker Cecil
Dionex Dionex / Gynkotek GBC GE AKTA Gilson
ICI Jasco Jenway Knauer Kontron
LDC LKB and Pharmacia Merck Hitachi Perkin Elmer (D2) Perkin Elmer (HC)
Shimadzu Thermo Unicam (D2) Unicam (HC) Varian (D2)
Varian (HC) Waters
  Request an engineer call out Call 01480 217181 - or - Fill in the form below: [contact-form-7 id="4" title="Log a service call"]